Agilent Technologies 8753ET User Manual

Page 9

Advertising
background image

9

1. One-port or enhanced-response calibration.
2. Enhanced-response or response-only calibration.
3. 2 dB less with Option 004.
4. Option 004 may degrade uncorrected source

match as much as 2 dB.

Transmission uncertainty
(enhanced-response calibration)

Transmission uncertainty
(response calibration)

Reflection uncertainty
(one-port calibration)

Agilent 8753ET (50-ohm systems)

Type-N test ports

Configuration
Network analyzer

8753ET Standard,
Option 006, or Option 004

Calibration kit

85032B

Test port cable

Part number 8120-4781

System dynamic range
300 kHz to 16 MHz

100 dB

16 MHz to 3 GHz

110 dB

3 GHz to 6 GHz

105 dB

Measurement Port Characteristics

Frequency Range

Uncorrected

300 kHz–1.3 GHz 1.3 GHz–3.0 GHz

3.0 GHz–6.0 GHz

Directivity

30 dB

24 dB

19 dB

Source match

4

25 dB

20 dB

14 dB

Load match

24 dB

19 dB

16 dB

Reflection

±1.0 dB

±1.0 dB

±2.0 dB

tracking

Transmission

±1.5 dB

±1.5 dB

±2.5 dB

tracking

Crosstalk

100 dB

100 dB

90 dB

Measurement Port Characteristics (continued)

Corrected

Reflection
measurements

1

300 kHz–1.3 GHz 1.3 GHz–3.0 GHz

3.0 GHz–6.0 GHz

Directivity

50 dB

47 dB

40 dB

Source match

42 dB

36 dB

31 dB

Load match
One-port cal

24 dB

19 dB

16 dB

Enhanced-

24 dB

19 dB

16 dB

reflection cal

Reflection

±(0.009 dB

±(0.019 dB

±(0.07 dB

tracking

+0.01 dB/°C)

+0.02 dB/°C)

+0.03 dB/°C)

Transmission
measurements

2

Source match
Enhanced-

42 dB

36 dB

31 dB

response cal

Response-only cal

3

25 dB

20 dB

14 dB

Transmission tracking
Enhanced-

±(0.006 dB

±(0.018 dB

±(0.054 dB

response cal

+0.01 dB/°C)

+0.02 dB/°C)

+0.03 dB/°C)

Response-only cal ±(0.033 dB

±(0.1 dB

±(0.27 dB

+0.01 dB/°C)

+0.02 dB/°C)

+0.03 dB/°C)

Advertising