External source mode

1

(CW time sweep only)

Frequency range

300 kHz to 6 GHz

R-channel input requirements

2

Power level

0 to -25 dBm

Spectral purity
Maximum spurious input <-30 dBc
Residual FM

<20 kHz

Typical settling time

500 ms (automatic)
50 ms (manual)

Frequency readout accuracy

2

0.1% (automatic)

Input frequency accuracy requirement

2

Manual

-0.5 to 5 MHz

Accuracy

(See magnitude and phase
characteristics)

Magnitude characteristics

Dynamic accuracy

(10 Hz IF BW)

Display resolution

0.001 dB/division

Marker resolution

3

0.001 dB

Trace noise
(0 dBm at transmission port or +5 dBm at reflection port,
ratio measurement, 3 kHz BW)

300 kHz to 3 GHz

<0.006 dB rms

3 GHz to 6 GHz

<0.010 dB rms

Reference level

Range

±500 dB

Resolution

0.001 dB

Stability

2

300 kHz to 3 GHz

0.02 dB/1° C

3 GHz to 6 GHz

0.04 dB/1° C

Phase Characteristics

Dynamic accuracy (10 Hz IF BW)
Range

±180°

Display resolution

0.01°/division

Marker resolution

3

0.01°

Trace noise
(0 dBm at transmission port or +5 dBm at reflection
port, ratio measurement, 3 kHz BW)
300 kHz to 3 GHz

<0.038° rms

3 GHz to 6 GHz

<0.070° rms

Reference level

Range

-180° to +180°

Resolution

0.01°

Stability

30 kHz to 3 GHz

0.05°/° C

3 GHz to 6 GHz

0.20°/° C

Polar characteristics

Range

10 x 10

–12

to 1000 units full

scale

Reference

±500 units

1. See the Agilent 8753ET/ES descriptions and options for a functional description.

Measurement accuracy is dependent on the stability of the input signal.

2. Typical performance.
3. Marker resolution for magnitude, phase and delay is dependent upon measured

value. Resolution is limited to five digits.

Advertising
This manual is related to the following products: