The WRM is capable of reading two static resistance values at the same time, and
their values are referred to as “R1” and “R2” on the graph. The LTCA software can
also calculate each resistance’s value at a reference temperature. These values are
referred to as “R1s” and “R2s”, respectively.

LTCA-10 and LTCA-40

In static resistance test mode, the LTCA is capable of reading three static resistance
values at the same time, and their values are referred to as “R1”, “R2”, and “R3” on
the graph. The LTCA software can also calculate each resistance’s value at a
reference temperature. These values are referred to as “R1s”, “R2s”, and “R3s”,
respectively.

In dynamic resistance test mode, only the “R1” resistance values are recorded over
time. For dynamic resistance tests, the LTCA software displays the resistance curve
for “R1”, the test current graph for “R1” (referred to as “Test Current”), and the LTC
motor current graph (referred to as “Motor Current”). The colors for each graph can
be customized.