UMC 600IS Installation Manual

Parameter

Description

Values

1

Graduations

NTEP to 10,000 (80,000 available).

2

Display resolution

1, 2, 5, 10, 20, 50, 100

3

Decimal point

0, 0.0, 0.00, 0.000, 0.0000

4

Span select

Parameter 4 is not configurable: Span selection is automatically set for the
following ranges: Lo = 1.7 – 3.3 mV/V; Hi = 0.5 – 2.0 mV/V

5

Digital averaging

1, 2, 4, 8, 16, 32, A1 = 8-4-2; A2 = 16-8-2

6

Tare mode

ATNR, AUTO, FIXED, BOTH (inhibit with motion)

7

AZM band

Off, 0.5, 1, 3, 5, and 10 divisions
Use 0.5 for H-44, bench, counter and livestock applications; use 3 for vehicle,
axle-load, and railroad scales

8

AZM/PAZ aperture

±1.9%, 100% of capacity includes push-to-zero, H-44: 1.9%

9

Motion

Off, 1, 3, 5, 10, 20, 50 divisions H-44: vehicle, axle, livestock, RR, 3.0; all other 1.0

10

Displayed units

Lb and kg conversion

11

Setpoint mode

Not supported

12

Zero band

Not supported

13

Weigh mode

Not supported

14

Serial output

Port 1, demand, continuous, baud rate, G/T/N, or display (Port 2 not supported)

Table 3-1. Parameter Overview

Figure 3-4. Configuration Menu

XXXXXXX

XXXXXXX

XXXXXXX

GRADS

1

5

2

10

0.0

0.00

OPTION 5

OPTION 4

OPTION 3

OPTION 1

OPTION 2

SERIAL

CALIBR

CONFIG

RES

resolution

DECPT

decimal point

50

20

100

OPTION 6

100

selection

list,

see table 3-2

on page 15

0.000

0.0000

1

4

2

8

AVERAGE

digital average

A1=8-4-2

16

32

ATNR

AUTO

A2=16-8-2

FIXED

BOTH

OFF

0.5

1

3

5

10

1.9

FS

TARE

enable tare

AZM

band

AZM/PAZ

band

1

Off

3

10

5

20

kg

lb

lb. conversion

MOTION

band

DISPLAY

50

SETPOINT

MODE

Not

supported

ZERO BAND

MODE

OPTION 7

WEIGHING

MODE

SERIAL

serial output

port 1

demand

continous

baud rate

G/T/N

display

Not

supported

Not

supported