0 appendix, 1 error and status messages – Rice Lake Condec UMC600IS User Manual
Page 39
36
UMC 600IS Installation Manual
6.0
Appendix
6.1
Error and Status Messages
The UMC 600IS indicator provides a number of error and status messages. When an error occurs, the message is
shown on the indicator display. Table 6-1 lists these messages and their meanings.
Error Message
Description
Solution
Err1
Loss of configuration
Reconfigure
Err1a
Loss of option configuration
Err1b
Err1c
Err2
Loss of zero calibration
Recalibrate
Err3
Loss of span calibration
Err5
Loss of auto tare
Acquire tare
Err6
Loss of auto zero
Acquire zero
Err9
Not in span calibration
Program span value before entering
Err 10H
Input > 3.3 mV/V at full scale
Recalibrate span to set full scale in the range 0.4–3.3 mV/V
Err 10L
Input < 0.4 mV/V at full scale
Err 10P
PTZ (push to zero) > 2%
Perform zero calibration
Err 10–
PTZ > weight used for span calibration
OL
Gross overload
Check load cell wiring
UL
Gross underrange
CAL
Calibration successful
—
Table 6-1. UMC 600IS Error and Status Messages