Configuration

23

3.3.4

Option 4 – Auto Shutoff Mode

The auto shutoff mode can prolong battery life for
applications using the battery input power to the
indicator.

Option 4 allows the indicator to be configured to shut
off automatically when idle for a specified time (1–9
minutes). Parameter 4.2 can be enabled to reset the
shutoff timer if motion is detected on the scale after
the timer is started.

Option 4 is not a “sleep mode”; once the indicator
shuts off, it must be turned on again. With Option 4
enabled, the on/off switch becomes active (see
Figure 3-12) and can be used to restart the unit.

NOTES:

Option 4 requires hardware setup inside the
indicator enclosure and must be enabled at the
factory.

Option 3 (time and date) must be enabled and
set to use Option 4.

Prompts for auto shutoff mode (OP.4) are shown in
Table 3-13.

Figure 3-12. On/Off Switch for Option 4

3.3.5

Option 6 – Five-Point Linear Calibration

There are five calibration points that can be entered
sequentially. Calibrations using fewer than five points
will linearize the curve up to the last data point.

E n a b l i n g O p t i o n 6 p l a c e s t h e i n d i c a t o r i n a
linearization calibration when the unit is in the
calibration mode (SW1-1 closed on front panel DIP
switch). Use the five-point linearization procedure
outlined in Section 4.3 on page 30 to perform a scale
calibration when Option 6 is enabled.

The five-point linearization calibration (OP.6) is
shown in Table 3-14.

Prompt Display

Interpretation

Auto Shutoff Mode

OP.4

oFF

Off: Auto shutoff mode disabled

OP.4

on

On: Auto shutoff mode enabled

Shutoff timer value

4.1

n

Set number of minutes, 1–9, before
shutoff

Motion detection

4.2

on

On: Shutoff timer is reset if motion is
detected before timer expires

4.2

oFF

Off: Indicator always shuts down when
shutoff timer expires

Table 3-13. Option 4 Configuration Selections

1

6

2

7

3

8

4

9

5

0

CE

ENT

TARE

RECALL

ZERO

NET

GROSS

TARE

lb/kg

CONV

PRINT

1 SET

POINT

2 SET

POINT

CAPACITY

600 x 0.1

ZERO

NET

GROSS

lb

kg

MOTION

ON/OFF switch

Prompt

Display

Interpretation

Notes

Five-point linearization

OP.6

oFF

Off

OP.6

on

On

Enter up to five linearization
points

Table 3-14. Option 6 Configuration Selections