About this manual, 0 introduction, Warning – Rice Lake Condec UMC600IS User Manual

Page 4: 1 factory mutual approval

Advertising
background image

Introduction

1

About This Manual

The UMC 600IS digital weight indicator represents
t h e l a t e s t i n s t a t e - o f - t h e - a r t m i c r o p r o c e s s o r
technology specifically applied to the weighing
marketplace. This manual provides information on
installation, calibration, configuration, and operation
of the UMC 600IS.

The installer should be familiar with the National
Electrical Code and RP 12.6 (

Recommended Practice

)

requirements for installation of equipment in
hazardous areas (NEC Article 504,

Intrinsically Safe

Systems

) published through the Instrument Society of

America.

The UMC 600IS is a Factory
Mutual-approved system. This
approval is valid only if the
installation conforms to the
guidelines described in this manual. If

modifications are made to the installation procedure, the
enclosure is opened, or the instrumentation is changed in
any way, Factory Mutual approval is void and the
customer becomes fully responsible for such
modifications.

This instrument and accompanying equipment must be
installed and serviced by an authorized technician in
accordance with the instructions provided in this manual.
Improper specification, installation, or service of this
equipment could result in personal injury or property
damage.

Authorized distributors and their
employees can view or download this
manual from the Condec distributor
site at

www.4condec.com

.

1.0

Introduction

The UMC 600IS is a single-channel digital weight indicator designed and approved to operate as an intrinsically
safe system in a wide variety of scale and weighing applications. The indicator is housed in a NEMA 4X
stainless steel sealed case. The standard unit is equipped with a tilt stand base for tabletop or wall mounting
applications. The indicator front panel consists of a 21-button keypad, six digit, seven-segment display and 11
LED annunciators (see Figure 1-1 on page 2).

Features of the UMC 600IS include:

Full front panel digital configuration and
calibration

Zero and span temperature compensation to
ensure compliance with NTEP temperature
range requirements (–10 to 40°C)

Nonvolatile memory stores data for
calibration, temperature compensation,
configuration parameters, auto or fixed tare
values, and PAZ and AZM values

T

en-thousand displayed graduations;

80,000 graduation expansion available

NOTE:

Use of more than 20,000 graduations may cause

undesirable display instability in some applications.

Analog sensitivity to 0.3

µ

V/grad at 20,000

graduations

Ten updates per second, with selectable
digital averaging and auto averaging; 5 Hz
active analog filter for smooth response

Excitation for four 350

load cells at 5 VDC

Operable from DC power supply or optional
battery

Time and date print selection

1.1

Factory Mutual Approval

The UMC 600IS is Factory Mutual (FM) approved
for:

Classes I, II, and III,

Divisions 1 and 2, and

Groups A, B, C, D, E, F and G

The Factory Mutual (FM) approval is a system
approval for all devices connected directly to the
UMC 600IS digital weight indicator.

Other devices that have FM Entity Approval cannot be
used with this system unless specifically listed in this
manual as part of the Condec Factory Mutual systems
approval. Failure to comply with this voids the FM
approval. The control drawing is the guide to the
installation of this system.