Rice Lake Condec UMC600IS User Manual

Page 17

Advertising
background image

14

UMC 600IS Installation Manual

Prompt Display

Interpretation

Notes

Digital Averaging

No.

Averages

Update

Rate

5

1

1

10/sec

5

2

2

5/sec

5

4

4

2.5/sec

5

8

8

1 sec

5

16

16

2 sec

5

32

32

4 sec

5

A1

8-4-2

Variable

5

A2

16-8-2

Variable

Tare Enable

Selection of either fixed tare (Ft) or fixed tare and auto tare (both) in Parameter 6
allows up to a six digit fixed tare entry to be made using the numeric front panel
keys or, when configured for full duplex serial communication, a fixed tare entry
can be down loaded through the serial port.

6

Atnr

Auto (stored) tare only
– no recall

6

AUto

Auto (stored) tare only

6

Ft

Fixed (manual) tare
only

6

both

Auto or fixed tare

AZM Capture Band (Displayed Grads)

7

oFF

Off

7

0.5

±0.5

7

1

±1.0

7

3

±3.0

7

5

±5.0

7

10

±10.0

Table 3-4. Configuration Display Prompts 5 – 7