Chapter 2

_________________________________________________________________________

________________________________________________________________________

BE6027-12

LumaSoft Gas Software

LumaSense Technologies A/S

7810 and 7860

Page 20 of 194

Baud rate

9600

Stop bits

1

Data bits

7

Parity

Even

Hardwire mode

Leased line

Handshake type

Hardwire

Table 2.2 Monitor RS232 communication parameters

These are set as the default values in the Monitor.

To prevent communication errors, the text line terminator, print data log

and print error log must be set as shown in Table 2.3.

Text line Terminator CR-LF

Print Data Log

NO

Print Error Log

NO

Table 2.3 More Monitor parameters

2.2.7

Checking/Changing the RS-232

Communication

Parameters in the Monitor

The communication parameters for the serial interface must be set using

the push-buttons on the front of the Monitor.

1. Press SET-UP S3 S1 S3 S1 (see

Chapter 1.3

) .The screen display

shows the following text.

If the baud rate displayed is incorrect press

and use ▲ and ▼ to

display the correct value. Press

again to store the selection.

If the baud rate displayed is correct, then press ▼ to continue to the

next parameter.

Press S1 to select 1 STOP BIT.

Press S1 to select 7 DATA BITS

Press S2 to select EVEN PARITY

Press S3 to select LEASED-LINE

Press S3 to select HARD-WIRED HANDSHAKE.

Press SET-UP to exit the set-up mode

Press RESET and S1 in order that the new settings are enabled.