LumaSense Technologies LumaSoft Gas Multi Point 7860 User Manual

Page 39

Advertising
background image

Chapter 3

_________________________________________________________________________

________________________________________________________________________

BE6027-12

LumaSoft Gas Software

LumaSense Technologies A/S

7810 and 7860

Page 39 of 194

By selecting Auto the monitoring system controls the flushing by

selecting the optimum flushing time. This time can vary as the option

takes into account the cleanness of the external air filters and the length

of the sampling tube. Type in the length of the tubing. (See

Figure

3.18

).

Selecting Fixed Time allows the user to define a flushing time for

chamber and tubing. Table 3.1 shows the acceptable values for flushing

chamber and tubing.

Acceptable Values

Interval

Default

Value

Chamber: 2 to 60 s

Tube: 0 (=off), 3 to 120 s

1 s

1 s

8 s

3 s

Table 3.1 Acceptable values for flushing the system when selecting Fixed

Time flushing

In the Compensation group box, see Figure 3.18, compensation for

Water and Cross Interference can be checked on/off.

It is suggested that Water should be compensated for in all measuring

situations except where extremely dry gases are being measured.

If any interfering gas other than Water Vapour is present in the ambient

air sample, the Monitor can cross compensate for the interference

caused by its presence provided that a selective optical filter is installed

in the Monitor to measure the concentration of the interfering gas

present.

In the Sample Information group box, see Figure 3.18, select whether

multiplexer sequences are continuous. Select Sample Continously if

you want the next multiplexer sequence immediately after finishing the

previous multiplexer sequence. If Sequence Interval is selected a time

interval (in minutes) can be set to define the start of one multiplexer

sequence to the beginning of the next. If no multiplexer is used the time

interval will be between the individual samples.

Gas Index Card

Select the Gas index card and press the Read Filter Info soft-key (

See

Figure 3.19

) to read the filter configuration from the Monitor.