5 units – LumaSense Technologies LumaSoft Gas Multi Point 7860 User Manual

Page 49

Advertising
background image

Chapter 3

_________________________________________________________________________

________________________________________________________________________

BE6027-12

LumaSoft Gas Software

LumaSense Technologies A/S

7810 and 7860

Page 49 of 194

3.2.5 Units

Select Unit in the Set Up drop down menu. (See Figure 3.31)

Figure 3.31 Set Up pull-down: Unit

Figure 3.32 Select in which units the data should be displayed

In this window the user can select in which units Humidity, Pressure,

Gas Concentration, Temperature and Length will be shown. The

Normalization Temperature can also be changed. The normalization

temperature is the reference temperature, for reporting data in absolute

unit mg/m

3

. A normalization temperature of 25°C means that the

measured gas concentrations will be calculated in mg/m

3

at 25°C. The

user can choose between three different normalization temperatures,

see table 3.2.

Display Text

Options

Default

Value

NORMALIZATION TEMPERATURE

0°C, 20°C and 25°C

20°C

NORMALIZATION TEMPERATURE

32°F, 68°F and 77°F

68°F

NORMALIZATION TEMPERATURE

273K, 293K and 298K

293K

Table 3.2 The possible values and the default values of the

normalization temperature