Modbus Communication

25

Modbus/TCP

The Modbus/TCP protocol is used on Ethernet communication. The XTDL-HT™ supports traditional IPv4
internet layer and EUI-48 MAC addressing. Each Ethernet device on a local network must have unique
address. If multiple XTDL-HT™ instruments are installed, each unit must be configured to a unique MAC
address and a unique IP addresses. New address goes into effect once power to the instrument is
recycled.

Change Ethernet IP Address

Go to Communication

>

Ethernet

>

Address

Change Ethernet subnet mask

Go to Communication

>

Ethernet

>

Subnet Mask

Change Ethernet default gateway address

Go to Communication

>

Ethernet

>

Gateway

Change the wait time on Ethernet port for a
complete command message from the master

Go to Communication

>

Ethernet

>

Timeout

Change Modbus Ethernet port

3

Go to Communication

>

Ethernet

>

Port

Change Ethernet MAC address

Go to Communication

>

Ethernet

>

MAC

Change Modbus address on Ethernet port

Go to Communication

>

Modbus

>

Ethernet

>

Address



3

Modbus/TCP uses port 502. User is not recommended to change port assignment unless the Modbus master is

known to use a different port.