COSA Xentaur Xentaur XDTL-HT User Manual

Page 46

Advertising
background image

46

Register
Number

Address

Access

Level

Data
Type

Description

Values

40305

304 (130h)

Admin

16-bit

Relay 1 mode

bit 0: Polarity (0=normally closed, 1=normally
opened)
bits 1-8: Variable ID

4

bit 9: Assert when variable below low threshold
(0=disabled, 1=enabled)
bit 10: Assert when variable above high
threshold (0=disabled, 1=enabled)
bit 11: Self cleared (0=disabled, 1=enabled)
bit 12: Unused
bit 13: Assert on fault (0=disabled, 1=enabled)
bit 14: Time controlled (0=disabled, 1=enabled)
bit 15: Remote controlled (0=disabled,
1=enabled)

40306

305 (131h)

Admin

16-bit

Relay 2 mode

See Relay 1 mode

40307

306 (132h)

Admin

16-bit

Relay 3 mode

See Relay 1 mode

40308

307 (133h)

Admin

16-bit

Relay 4 mode

See Relay 1 mode

4

Valid relay variable ID’s are:

Value

Description

Value

Description

Value

Description

Value

Description

0

Off

5

Analog output 1 valid

67

Enclosure temperature

120

PID 1

1

System fault

6

Analog output 2 valid

68

User input 1

121

PID 2

2

Calibration fault

64

Dewpoint

69

User input 2

122

PID 3

3

Calibrating

65

Sample temperature

70

User input 3

123

PID 4

4

Validating

66

Sample pressure

71

User input 4

Advertising