COSA Xentaur Xentaur XDTL-HT User Manual

Page 47

Advertising
background image

47

Appendix B: Modbus Registers

Register
Number

Address

Access

Level

Data
Type

Description

Values

Digital input registers

40351

350 (15Eh)

Admin

16-bit

Digital input 1 mode

bit 0: Polarity (0=active high, 1=active low)
bits 1-7: input variable ID
bits 8-15: Unused

40352

351 (15Fh)

Admin

16-bit

Digital input 2 mode

bit 0: Polarity (0=active high, 1=active low)
bits 1-7: input variable ID
bits 8-15: Unused

40353

352 (160h)

Admin

16-bit

Digital input 3 mode

bit 0: Polarity (0=active high, 1=active low)
bits 1-7: input variable ID
bits 8-15: Unused

40354

353 (161h)

Admin

16-bit

Digital input 4 mode

bit 0: Polarity (0=active high, 1=active low)
bits 1-7: input variable ID
bits 8-15: Unused

Analog output registers

40401

400 (190h)

Read-Only

16-bit

Analog output 1 type

0=0-5V
1=0-10V
2=±5V
3=±10V
5=4-20mA
6=0-20mA
7=0-24mA

40402

401 (191h)

Admin

16-bit

Analog output 1 low limit DAC value

Valid values are 0 - 65535

40403

402 (192h)

Admin

16-bit

Analog output 1 high limit DAC value

Valid values are 0 - 65535

40404

403 (193h)

Admin

16-bit

Analog output 1 during calibration
operation

0=Set to low state
1=Hold output
2=Track calibration measurement

Advertising