Setup menu – COSA Xentaur Xentaur XDTL-HT User Manual

Page 30

Advertising
background image

30

Setup Menu

Variable
Alarm State

On Error
Self Cleared

Mode
Day

Time On
Time Off

Time

Channel 1

Variable
Alarm State

On Error
Self Cleared

Variable
Alarm State

On Error
Self Cleared

Variable
Alarm State

On Error
Self Cleared

Channel 2

Channel 3

Channel 4

Relays

Channel 3 and 4 menu would not appear if the instrument is not configured with the extra relays.

Set relay X to operate on a measurement Y alarm

Go to Measurement

>

Y

>

Low Alarm

Set the alarm low threshold
Go to Measurement

>

Y

>

High Alarm

Set the alarm high threshold
Go to Relays

>

Channel X

>

Variable

Select the desired measurement
Go to Relays

>

Channel X

>

Alarm State

Select the desired alarm condition

Set relay X to operate on a system event

Go to Relays

>

Channel X

>

Variable

Select the desired event
Go to Relays

>

Channel X

>

Alarm State

Select the relay polarity

Select how relay X behaves on system fault

Go to Relays

>

Channel X

>

On Error

Turn on/off self-cleared function on relay X

Go to Relays

>

Channel X

>

Self Cleared

NOTE:

Relay 2 is reserved for controlling the enclosure heater.