COSA Xentaur Xentaur XDTL-HT User Manual

Page 48

Advertising
background image

48

Register
Number

Address

Access

Level

Data
Type

Description

Values

40405

404 (194h)

Admin

16-bit

Analog output 1 variable

0=Off
64=Dewpoint
65=Sample temperature
66=Sample pressure
67=Enclosure pressure
68=User input 1
69=User input 2
70=User input 3
71=User input 4
120=PID 1
121=PID 2
122=PID 3
123=PID 4
32768=Remote controlled

40406

405 (195h)

Admin

Float

Analog output 1 low limit

Valid values are -100 - 99999.9

40408

407 (197h)

Admin

Float

Analog output 1 high limit

Valid values are -100 - 99999.9

40410

409 (199h)

Admin

16-bit

Analog output 1 remote controlled
value

Valid values are 0 - 65535

40411

410 (19Ah)

Read-Only

16-bit

Analog output 2 type

0=0-5V
1=0-10V
2=±5V
3=±10V
5=4-20mA
6=0-20mA
7=0-24mA

40412

411 (19Bh)

Admin

16-bit

Analog output 2 low limit DAC value

Valid values are 0 - 65535

40413

412 (19Ch)

Admin

16-bit

Analog output 2 high limit DAC value

Valid values are 0 - 65535

40414

413 (19Dh)

Admin

16-bit

Analog output 2 during calibration
operation

0=Set to low state
1=Hold output
2=Track calibration measurement