Maintenance, Warranty, Maintenance policy – COSA Xentaur Xentaur XDTL-HT User Manual

Page 37

Advertising
background image

Maintenance

37

Maintenance

Although the XTDL-HT™ laser cell is designed to withstand
high level of contamination, the instrument is
recommended for periodic cleaning to prevent
contamination buildup on the gas sampling line. Annual
inspection and cleaning are sufficient for general use, more
often for dirty application. Maintenance procedure
includes:

1. Check the pressure regulators and pressure gauge for

correct pressure. Remove and clean if necessary.

2. Check the flow meter for correct flow rate. Remove and

clean if necessary.

3. Flush the sampling system with solvent appropriate for

the application.

4. Remove bottom of the laser cell and drain any liquid.

Clean before re-installing.

Warranty

This instrument is warranted to be free from defects in workmanship and materials. Liability under this
warranty is limited to servicing, calibrating, and replacing any defective parts of the instrument
returned to the factory for that purpose. Fuses are specifically excluded from any liability. This
warranty is effective for one year from the date of delivery to the original purchaser. The equipment
must be examined by COSA Xentaur and found to have been defective for this warranty to be
considered valid.

If any damage is determined to have been caused by misuse or abnormal conditions of operation, the
owner will be notified and repairs will be billed at standard rates after approval.

Maintenance Policy

In cases when equipment fault is suspected, please notify your representative of the problem; be sure
to provide them with model and serial numbers. If the problem cannot be resolved, then ask for a
Return Material Authorization (RMA) number and shipping instructions. Issuance of an RMA number
does not automatically imply that the equipment is covered by our warranty that will be determined
after we receive the equipment. Pack the equipment in a suitable box with sufficient padding, include
the RMA number on your paperwork, and send the equipment, prepaid, to the designated address.
Equipment returned without an RMA number, or with reversed shipping or import/export charges, will
not be accepted.

Removing bottom of laser cell

1. Remove

duct cover to

expose cables

2. Remove

Swagelok

connectors

3. Twist off laser

cell bottom