HT instruments MACROTEST 5035 User Manual

Page 42

Advertising
background image

MACROTEST

5035

EN - 41

4.5. R

a15mA

: GLOBAL EARTH RESISTANCE MEASUREMENT


Turn the

switch on R

a15mA

position.

Switch on the instrument.



1. Connect the three black, blue and green connectors of the shuko cable or of the untied

cables in the corresponding input terminals of the instrument

B1, B3, B4 (see possible

connections in the following pictures). When using untied cables connect the crocodiles to
the free ends of the cables.

B2

B3

B4

B1

P

Connection for Global Earth Resistance in a 230V

single- or two-phase system

B2

B3

B4

B1

L1
L2

L3
N

Connection for Global Earth Resistance in a 400V

three-phase system with neutral

B2

B3

B4

B1

L1
L2

L3

Global Earth Resistance in a 230V or 400V three-phase system without neutral


2. Connect the shuko plug into a 230V 50Hz socket or the crocodiles to the conductors of

the three-phase system (see previous picture).