KYORITSU 3127 User Manual

Page 10

Advertising
background image

- 7 -

<Voltage monitor for insulation resistance range>

Rated voltage

250V

500V

1000V

2500V

5000V

Accuracy

Measuring

range

30-330V 30-650V 30-1200V 30-3000V 30-6000V ±10%rdg±20V

(resolution 10V)

This monitor is used to check whether electrical charge stored on the equipment

under test is discharged or not. The measured voltage value displayed on the

LCD is a reference value.

Please be noted that the indicated value, when external AC voltage is applied to

the instrument, is not the correct value.

<Volt meter >

Volt

Measuring range

Resolution

Accuracy

DC voltage:±30 - ±600V

AC voltage: 30 - 600V

(50/60Hz)

1V

±2%rdg±3dgt

Frequency

45.0-65.0Hz(*1)

0.1Hz

±0.2Hz

(*1)When the measured voltage is less than 30V or DCV is measured, the

frequency is displayed as “----Hz” on the LCD.

<Ammeter (Output Current)>

Measuring range

Accuracy

0.00nA~5.50mA

±10%rdg(*1)

(*1) If result of resistance measurement is 10MΩ or more, output current is

determined by resistance and voltage.(Accuracy is derived from measured

resistance specification and measured voltage specification)

<Capacity meter>

Rated voltage

Measuring range

Accuracy

250V ∼ 2500V range

5.0nF ∼ 50.0μF

±5%rdg±5dgt

5000V range

5.0nF ∼ 25.0μF

Calculated Value

PI,DAR,DD

Measurement Mode

Display range

Computational error

PI

0.00 ∼ 9.99

±2dgt

DAR

0.00 ∼ 9.99

±2dgt

DD

0.00 ∼ 9.99

±2dgt

Advertising