6-3 About BREAKDOWN mode and BURN mode
Either Breakdown mode or Burn mode can be set for each measurement: IR

PI/ DAR

,

SV, DD, RAMP.

(1) Breakdown mode

When applied voltage decreases drastically due to breakdown or current sudden

increase caused by insulation degradation, KEW3127 stops measurement
automatically so as not to develop further damage.

KEW3127 continues measurement until an event of breakdown or current

sudden increase is detected.

(2) Burn mode

When Buren mode is selected, KEW3127 continues measurement even if

breakdown or current sudden increase event occurs. Keep checking the
damage degree of the breakdown point and fi nd the insulation failure point.
Testing performed in Burn mode is destructive.

When 250V Range has been selected, it is unable to select the Breakdown

mode.

When breakdown occurs

and measurement stops,

breakdown mark blinks.