KYORITSU 3127 User Manual

Page 11

Advertising
background image

- 8 -

Display:

Liquid crystal display

Bar graph: Max. 41points

DAR/PI value: 9.99

Time: 99:59

Low battery warning: Battery mark display (in 4 levels)

Overrange indication: “OL” mark appears at insulation resistance

range and “Hi” mark at voltage range.

Auto-ranging:

Range shifts to upper range : 1000 counts

Range shifts to lower range : 80 counts

(Only at the insulation resistance range)

Auto-power-off:

Power off function operates in 10 min. after

the last switch operation.

Used location:

altitude 2000 m or less

Temperature & humidity range (guaranteed accuracy):

23ºC±5ºC/Relative humidity 85% or less

(no condensation)

Operating temperature & humidity range:

0ºC to 40ºC/Relative humidity 85% or less

(no condensation)

Storage temperature & humidity range:

-20ºC to +60ºC/Relative humidity 75% or less

(no condensation)

Overload protection: Insulation resistance range: AC1200V/10sec.

Voltage range: AC720V/10sec.

Withstand voltage:

AC6720V(50/60Hz)/5sec.

(Between electrical circuit and enclosure)

Insulation resistance: 1000MΩ or more/DC1000V

(Between electrical circuit and enclosure)

Dimension:

208(L)×225(W)×130(D)mm

(Hard case 380(L)×430(W)×154(D)mm)

Weight:

4kg approx. (including battery)

Total 8kg approx. (including accessories)

Power source:

Rechargeable battery

12V5Ah Lead acid batteries (PXL12050) or equivalent.

Power Adapter

Output:DC15V(15VA)

for recharging

Input:100-240V 50/60Hz