— 15 —

Class I Test Flowchart

Start

#

CAUTION

• Short-circuit between L/N and PE is doubt when "no" and "OFF"

appears on the LCD in turns at the insulation resistance measurement

of the Extension leads test for Class I & II while DUT is on. Testing

should be immediately suspended.

• Follow the procedure described in 5.2-3 and do NULL setting before a

measurement.

• Alligator clip must make good contact with the enclosure of the DUT.

• When the terminal is open or the resistance value exceeds measuring

range, OL mark (over range display) appears on the LCD.

• Do not touch the device under test while testing is in progress. Since

a high voltage of 500V, user may get electrical shock.

. Protective conductor resistance.

RPE <

= 1Ω?

.Appliance switch test

(3). Insulation resistance between L/N and PE.

LnE >

= 1MΩ?

Light up in red

no” →“Con”→“value” will

be repeated on the LCD.

Light up in red

no” →“LnE”→“value” will

be repeated on the LCD.

Value (1). and (3). will be

alternately displayed on LCD

When the resistance between

LN is about 100kΩ or more

no” →“OFF” is displayed

alternately on the LCD.

No

No

No

Yes

Yes

Yes

PASS

Light up in green

Light up in green

Is the Switch

turning on?

Advertising