Fuse replacement – KYORITSU 6201A User Manual

Page 26

Advertising
background image

— 24 —

#

WARNING

• Operate a device and measure the leakage current flowing on it at

Leakage current test function. Care should be taken not to touch with

the heating or movable parts during tests. Extra care should also be

taken to light and heat generated by the device.

• Firmly insert the plugs of DUT to the AU socket of this instrument.

Plugs may be heated if Leakage current test is performed with

improper connection.

• Do not connect/remove the plugs during Leakage current test. It may

cause reading error. Do not use the instrument on the device which

has a power of 2kVA or more.

7. Fuse replacement


When the fuse blows during use, please replace with new one according to
below procedure.

Use a flat head screwdriver and turn it about 45º to left and remove the

fuse cap and fuse.

Remove the fuse from the fuse cap and replace it with new one.

Install the fuse cap and fuse again. At that point, the screwdriver groove

shall be at about 45º turned to left from the initial position. Use the flat
head screwdriver and turn it to right.

(The screwdriver groove will stop at the horizontal position.)

Fig. 15