Extension Leads Test  Flowchart

Start

#

CAUTION

• Short-circuit between L/N and PE is doubt when "no" and "OFF"

appears on the LCD in turns at the insulation resistance measurement

of the Extension leads test for Class I & II while DUT is on. Testing

should be immediately suspended.

• Follow the procedure described in 5.2-3 and do Null setting before a

measurement.

• When the terminal is open or the resistance value exceeds measuring

range, OL mark (over range display) appears on the LCD.

• Do not touch the device under test while testing is in progress. Since

a high voltage of 500V, user may get electrical shock.

(1). Protective conductor resistance test.

RPE <

= 1Ω?

(2). Insulation resistance test between L/N and PE.

(3). Polarity test between L-L and N-N.

L-L & N-N <

= 10Ω?

Light up in red

no” →“Con”→“value” will

be repeated on the LCD.

Light up in red

no” →“LnE”→“value” will

be repeated on the LCD.

Light up in red

no” →“L-L” →“value” or

no” →“n-n” →“value” will

be repeated on the LCD.

Value of (1) and (2) will be

alternately displayed on LCD.

No

No

No

Yes

Yes

Yes

PASS

Light up in green

Light up in green

LnE >

= 1MΩ?