KYORITSU 6201A User Manual

Page 19

Advertising
background image

— 17 —

#

CAUTION

• When the terminal is open or the resistance value exceeds measuring

range, OL mark (over range display) appears on the LCD.

• The test will stop. A message Stp appears on the LCD but values of

Leakage current aren't displayed.

#

WARNING

• Operate a device and measure the leakage current flowing on it at

Leakage current test function. Care should be taken not to touch with

the heating or movable parts during tests. Extra care should also be

taken to light and heat generated by the device.

• Firmly insert the plugs of DUT to the AU socket of this instrument.

Plugs may be heated if Leakage current test is performed with

improper connection.

• Do not connect/remove the plugs during Leakage current test. It may

cause reading error. Do not use the instrument on the device which

has a power of 2kVA or more.

6.3 Class Test
The Class Ⅱ appliances have the indication of “DOUBLE INSULATION”
or the symbol of . Double insulation test is to check the insulation
resistance of the appliances is within the range defined in the standards.

Fig. 13

Switch ON the power.

Advertising