— 19 —

Class II Test (Select the Leakage current test) Flowchart

Start

#

CAUTION

• When the terminal is open or the resistance value exceeds measuring

range, OL mark (over range display) appears on the LCD.

• The test will stop. A message Stp appears on the LCD but values of

Leakage current aren't displayed.

#

WARNING

• Operate a device and measure the leakage current flowing on it at

Leakage current test function. Care should be taken not to touch with

the heating or movable parts during tests. Extra care should also be

taken to light and heat generated by the device.

• Firmly insert the plugs of DUT to the AU socket of this instrument.

Plugs may be heated if Leakage current test is performed with

improper connection.

• Do not connect/remove the plugs during Leakage current test. It may

cause reading error. Do not use the instrument on the device which

has a power of 2kVA or more.

Power on Leakage current test.

LEA <

= 1mA ?

Light up in red

Stop the test and

no” →“LEA”→“value” will

be repeated on the LCD.

Value will be displayed on the LCD.

No

Yes

PASS

Light up in green

Brief indications“S-”and “on”will be displayed

in turns on the LCD for 5 sec and prompt the

user to set the DUT switch on.

Then a mark indicating working DUT function

will be displayed for 10 sec.

5sec

10sec