— 18 —

Class II Test Flowchart

Start

#

CAUTION

• Short-circuit between L/N and PE is doubt when "no" and "OFF"

appears on the LCD in turns at the insulation resistance measurement

of the Extension leads test for Class I & II while DUT is on. Testing

should be immediately suspended.

• When the terminal is open or the resistance value exceeds measuring

range, OL mark (over range display) appears on the LCD.

• Do not touch the device under test while testing is in progress.Since

a high voltage of 500V, user may get electrical shock.

6.4 ClassTest (Select the Leakage current test)

When the function is set to Class II test while pressing Select switch, Power

on leakage current test will be carried out instead of Insulation resistance

test. When Leakage current test is chosen, DUT will operate actually. Clip a

metal part of the machine tool except for the rotating part with Test Lead

M-7129A. In order to stop testing the leakage current while measuring, press

START/STOP . To start a test again, press START/STOP again after about

2 seconds or more from the stop. The test will be restart.

(1). Appliance switch test

Is the Switch

turning on?

(2). Insulation resistance between L/N and PE.

LnE >

= 1MΩ?

When the resistance between

LN is about 100kΩ or more

no” →“OFF” is displayed

alternately on the LCD.

Light up in red

no” →“LnE”→“value” will

be repeated on the LCD.

Value of (2). will be

displayed on the LCD.

No

No

Yes

Yes

PASS

Light up in green

Light up in green