— 16 —

6.2 Class I Test (Select the Leakage current test)
Selecting “Class I test” while Select Switch is being pressed down initiates
Leakage current test instead of Insulation resistance test.
In case of selecting Leakage current test, metal parts other than the heating
or movable parts must be clipped with Test Lead M-7129A since the DUT
activates. Pressing the START/STOP switch during Leakage current test
stops the test immediately. To restart the test, press the START/STOP

switch again 2 sec or later than the stop of the test. Then (1) Protective
conductor resistance will be restarted.
Fig 12 indicates how to connect the devices.

Class I Test (Select the Leakage current test) Flowchart

Start

(1). Protective conductor resistance.

RPE <

= 1Ω?

(2). Power on Leakage current test.

LEA <

= 5mA?

Light up in red

no” →“Con”→“value” will

be repeated on the LCD.

Light up in red

Stop the test and

no” →“LEA”→“value” will

be repeated on the LCD.

Value of (1). and (2). will be

alternately displayed on the LCD.

No

No

Yes

Yes

PASS

Light up in green

Light up in green

Brief indications“S-”and“on”will be displayed

in turns on the LCD for 5 sec and prompt the

user to set the DUT switch on.

Then a mark indicating working DUT function

will be displayed for 10 sec.

5sec

10sec