WARNING

• Operate a device and measure the leakage current flowing on it at

Leakage current test function. Care should be taken not to touch with

the heating or movable parts during tests. Extra care should also be

taken to light and heat generated by the device.

• Firmly insert the plugs of DUT to the AU socket of this instrument.

Plugs may be heated if Leakage current test is performed with

improper connection.

• Do not connect/remove the plugs during Leakage current test. It may

cause reading error. Do not use the instrument on the device which

has a power of 2kVA or more.

6.6 Leakage Current Test

This function is to conduct the leakage current test separately with DUT

operating actually. A leakage current only will be displayed on the LCD after

DUT is electrified for 15 seconds.

LED will not light on even the Class I, Class II thresholds are exceeded.

• Set the function switch to Leakage Current Test position.

• Refer to the Fig12 or Fig13 for connection of an appliance.

• After set up is done, press START/STOP switch.

• Check the switch of DUT is ON.

• DUT will operate for 15 seconds, and the maximum value of the leakage

current will be displayed on the LCD. If the leakage from the DUT is

greater than 20mA then the 6201A should immediately stop the test.

• In order to stop the Leakage Current Test, press START/STOP switch

again. The test will stop. A message “Stp” appears on the LCD but values

of Leakage current aren't displayed.

Leakage Current Measment Flowchart

START

LEA <

= 20mA?

Stop the test and “OL” will

be displayed on the LCD.

Value of leakage current will

be displayed on the LCD.

Yes

No

Brief indications "S-" and "on" will be

displayed in turns on the LCD for 5 sec and

prompt the user to set the DUT switch on.
Then a mark indicating working DUT function

will be displayed for 10 sec.

5sec

10sec