Page 11

Sonic-Pro

Industries, Ltd.

Industries, Ltd.

ProSeries

by Blue-White Ind.

TM

NO

NO

Down flows

must have

back pressure

Pipe must be full

Air could be trapped

!

3

Determine the
transducer mounting mode

DOPPLER

All pipe types and sizes

When operated in the Doppler mode, the transducers
are always mounted directly opposite each other, 180
degrees around the pipe.

When operated in the Transit-Time mode, the front
faces of the transducers must face each other and be
positioned the correct distance apart. In V-mode, the
transducers are on the same side of the pipe. In Z-
mode, the transducers are on the opposite side of the
pipe, 180 degrees apart.

When factory configured, the appropriate Transit-
Time
mounting mode is pre-configured and printed on
the serial label.

“V” Mount Mode

Typically smaller pipe

sizes from 2” to 6”

Separation

Distance

TRANSIT TIME

“Z” Mount Mode

Typically larger pipe

sizes from 4” to 100”

Separation

Distance

See section 6.2 for more information

Flow

or

Flow

Flow

or

Flow

Flow

or

Flow

TRANSDUCER MOUNTING LOCATION

NO

NO

Air bubbles may be present

OK

Sediment may be present

Horizontal

O

45

O

45

4

Determine the 1st
transducer location

The mounting surface must be
clean, smooth and free of
surface imperfections. Remove
all insulation material, loose
paint, etc. Clean the pipe
thoroughly. Use sandpaper if
necessary to remove surface
imperfections. Be sure to locate
the transducers on the side of
horizontal runs of pipe. Do not
mount the transducers over
weld seams. Place a mark
where the first transducer will
be located.

See section 6.5 for more information