Power On
and
Self-Test

LCD DISPLAY

On application of power, the power LED (green) on the motherboard comes on and the
flowmeter performs self-test functions. The approximate duration of self-test is 10
seconds. The flowmeter indicates that it is in the Self-Test State as follows:

Below is a diagram of the Sonic-Pro’s major operating states. The condition of the
motherboard LEDs is pictured under the name of each state. Transitions are annotated
with their causes, which may be user actions via the motherboard controls and timeouts
(uppercase text), user actions using a display optionally fitted to the flowmeter, (italic
text) or automatic transitions when a step is complete (in parentheses).

In the Self-Test state, the flowmeter checks the operation of internal circuitry, such as
communication between the processor and its peripheral functions, Verify that all LEDs
and the LCD display, if fitted, are working properly. Do not press any buttons during
self-test.

If an error condition is detected during self-test, the flowmeter enters the Faulted State.
Failure of the LEDs, LCD display, and the display touch-pad are not internally
detectable.

If self-test completes without error, the flowmeter enters the Startup State. Note that
completion of self-test does not guarantee that the flowmeter is properly configured and
able to make accurate measurements.

Model S1: the red and all four yellew LED
indicator lights flash at 1 Hz.

Model S2: as for Model S1, plus the LCD
shows the firmware version number and the
text “Self-Test ...”.

Model S3: as for Model S2.

INDICATOR LIGHTS IN

SELF-TEST STATE

Self-test …

Information Screen

Blue-White Industries, Sonic-Pro
Firmware Version, A01.00.00

TRANSDUCER POSITIONING

SELF-TEST

STARTUP STATE

ZERO CALIBRATION

RUN MODE

FAULTED STATE

SETUP MODE

Power ON

(fail)

(pass)

HOLD SETUP

FOR

3 SECONDS

(complete)

7 SECOND

TIME OUT

TAP SETUP BUTTON or

FLIP DOPPLER / TT

MODE SWITCH

TAP SETUP BUTTON or

TIME OUT (5 minutes) or

press “ESC” button

TAP SETUP

BUTTON

TAP SETUP
BUTTON or

Activate a

configuration

TAP SETUP
BUTTON or

press

“DONE” button

(fail)

(fail)

(fail)

SONIC-PRO OPERATING STATES

4.2
Description of
Operating
States and
Controls

4.0 Overview of Flowmeter Operation