Open the front cover of the SPU. Position the Doppler/Transit-Time switch to the
preferred operating mode (factory setting is transit time).

Operate the meter in the Doppler mode if the fluid to be measured contains

more than 10% particles.

The Sonic-Pro transducer connections are pre-wired at the factory. If a main AC input
power cord selection was made when ordering, the power cord and main AC power input
connection is factory wired. When powering the meter by DC, a power cable for this
purpose should be installed in one of the cable gland liquid-tight connectors. See section
3.8 below. The meter will automatically adjust for any main power input that is within
the power requirements range of 110 to 240 volts AC or 15 to 30 volts DC. See the
diagram below for wiring of output signals, communications signals and process control
relays.

The transducer cable length is fixed. Do not attempt to modify the length of the cables.
Various cable lengths are available from the factory. Contact the factory if you need
assistance. Shielded cable is recommended for signal output connections.

Operate the meter in the

Transit Time mode (best accuracy) if the fluid contains little or no particles (up to 10%
maximum).

3.6
Select the
Measurement
Method

Doppler/Transti-Time Switch

3.7
Electrical
Connections

CAUTION

risk of electric shock

All high voltage connections in the front wiring compartment are
considered “finger safe”. However,
use care when connecting high voltage wires to the plug-in type terminal
blocks.

disconnect electricity to the meter and

RS232

USB

ETHERNET

RS485

PROCESS CONTROL RELAYS

(THREE EACH) 10A 250V AC

R1

R2

R3

COM

NC
NO

COM

NC
NO

COM

NC
NO

TRANSDUCER

INPUTS

Tx 1

Tx 2

RED (+)

RED (+)

BLK (-)

BLK (-)

GRN

GRN

mA

Hz

POS (+)

NEG (-)

GRN

SIGNAL

OUTPUTS

NEG (-)

POS (+)

DC

AC

DC (+)

DC (-)

GRN

AC LINE

AC NTRL

POWER INPUT

SD CARD

DATA LOG
(REMOVE)

MODE SWITCH

T

D

INDICATOR

LIGHTS

COMMUNICATIONS

CONNECTOR PANEL

WARNING

INPUT POWER LIMITS

AC INPUT

110~240Vac 50/60Hz-40W

DC INPUT

15~30 Vdc-40W

SETUP BUTTON

1A 250V (5x20mm)

slow blow

littlefuse 239001.XP

FUSE

CAUTION

risk of electric shock

RED (+)

BLK (-)

GRN

WIRING CONNECTIONS

Page 18

Sonic-Pro

Industries, Ltd.

Industries, Ltd.

ProSeries

by Blue-White Ind.

TM

There are no user serviceable items inside the rear meter enclosure. If it
becomes necessary to open the back cover, be sure to disconnect the main
power supply before opening the back cover of the enclosure.

CAUTION

risk of electric shock