Proseries, Sonic-pro, 4 panel mounting 3.5 pipe mounting – Blue-White Sonic-Pro Hybrid Ultrasonic Flowmeters User Manual

Page 17

Advertising
background image

Page 17

Sonic-Pro

Industries, Ltd.

Industries, Ltd.

ProSeries

by Blue-White Ind.

TM

3.4
Panel
Mounting

3.5
Pipe
Mounting

7.90” (200mm)

1) The SPU can be panel mounted.

Measure and cut an opening in the
panel as shown.

2) Remove the two mounting plates from

the rear of the enclosure.

3) Install the gasket onto the rear of the

front panel of the SPU enclosure.

4) Place the enclosure into the panel.

5) Re-install the two mounting plates as

shown.

6) Thread the four 8-32 x 3.00” machine

screws through each of the two outer
tapped holes in the mounting plates.
Tighten the screws to secure the
enclosure to the panel.

10.40”

(264mm)

1) The SPU can be mounted on horizontal

or vertical pipe. The pipe must be secure
and of sufficient strength to support the
weight of the SPU.

2) Install the two mounting plates onto the

rear of the enclosure. For horizontal
pipe mounting, install the mounting
plates on the top and bottom plate cut-
outs. For vertical pipe mounting, install
the mounting plates on the side plate
cut-outs.

3) Thread the long mounting clamp

through the square cut-outs in the
mounting plates as shown.

4) Wrap the clamp around the pipe and

tighten loosely. Be sure the pipe is
seated in the pipe groove, then tighten
securely.

Panel

Mounting
Plates

Gasket

8-32 x 3.00”
Panel Screws

Advertising