Calibration tab, Samples, On 6.4 – PNI SeaTRAX User Manual

Page 32: 4 calibration tab

Advertising
background image

PNI Sensor Corporation

DOC#1018154 r02

SeaTRAX User Manual

Page 27

6.4

Calibration Tab

Note: The default settings are recommended for the highest accuracy and quality of calibration.

6.4.1

Samples

Before proceeding, refer to Section 5 for the recommended calibration procedure

corresponding to the calibration method selected on the Configuration tab.

Clicking the <Start> button begins the calibration process and immediately takes the first

sample.

If “Automatic Sampling” is not checked on the Configuration tab, it is necessary to click
the <Take Sample> button to take a calibration sample point. This should be repeated

until the total number of samples, as set on the Configuration tab, are taken while

changing the orientation of the module between samples as discussed in Section 5.

If “Automatic Sampling” is checked, the module will need to be held steady for a short
time and then a sample automatically will be taken. Once the window indicates the next

number, the module’s orientation should be changed and held steady for the next sample.

Once the pre-set number of samples has been taken (as set on the Configuration tab) the

calibration is complete.