7.3.23 kPowerUpDone (frame ID 23

d

)

This frame confirms the SeaTRAX received a command to power up. The SeaTRAX

will power up when it receives any signal on the native UART Rx line. The frame has no

payload. Since the module was previously powered down which drives the RS-232

driver TX line low (break signal), it is recommended to disregard the first byte.

7.3.24 kSetAcqParams (frame ID 24

d

)

This frame sets the sensor acquisition parameters in the SeaTRAX. The payload should

contain the following:

AcquisitionMode

This flag sets whether output will be presented in Continuous or Polled Acquisition

Mode. Continuous Mode is TRUE and is the default. Polled Mode should be

selected when the host system will poll the SeaTRAX for each data set. Continuous

Mode should be selected if the user will have the SeaTRAX output data to the host

system at a relatively fixed rate.

FlushFilter

Setting this flag to TRUE results in the FIR filter being flushed (cleared) after every

measurement. The default is FALSE.

Flushing the filter clears all tap values, thus purging old data. This can be useful if a

significant change in heading has occurred since the last reading, as the old heading

data would be in the filter. Once the taps are cleared, it is necessary to fully

repopulate the filter before data is output. For example, if 32 FIR taps is set, 32 new

samples must be taken before a reading will be output. The length of the delay before

outputting data is directly correlated to the number of FIR taps.

AcquireDelay

When operating in Continuous Acquisition Mode, the AcquireDelay sets the time

between samples taken by the module, in seconds. The default is 0.0 seconds, which

means the module will reacquire data immediately after the last acquisition. This is

an internal setting that is NOT tied to the time with which the module transmits data

to the host system. Generally speaking, the AcquireDelay is either set to 0, in which

case the SeaTRAX is constantly sampling, or set to equal the SampleDelay value.

AcquisitionMode

UInt8

FlushFilter

UInt8

Payload

AquireDelay

SampleDelay

Float32

Float32