Anritsu S251C User Manual

Page 25

Advertising
background image

FREQ/DIST

Displays the frequency and distance menu depending on the measurement mode.

Frequency
Menu

Provides for setting sweep frequency end points when

FREQ

mode is selected.

Selected frequency values may be changed using the keypad or Up/Down arrow
key.

q

F1

— Opens the F1 parameter for data entry. This is the start value for the

frequency sweep. Press

ENTER

when data entry is complete.

q

F2

— Opens the F2 parameter for data entry. This is the stop value for the

frequency sweep. Press

ENTER

when data entry is complete.

Distance
Menu

Provides for setting Distance to Fault parameters when a DTF mode is selected.
Choosing

DIST

causes the soft keys, below, to be displayed and the correspond-

ing values to be shown in the message area. Selected distance values may be
changed using the keypad or Up/Down arrow key.

q

D1

— Opens the start distance (D1) parameter for data entry. This is the start

value for the distance range (D1 default = 0). Press

ENTER

when data entry

is complete.

q

D2

— Opens the end distance (D2) parameter for data entry. This is the end

value for the distance range. Press

ENTER

when data entry is complete.

q

DTF AID

— Provides interactive help to optimize DTF set up parameters.

Use the Up/Down arrow key to select a parameter to edit. Press

ENTER

when

data entry is complete.

q

MORE

— Selects the Distance Sub-Menu, detailed below.

Distance
Sub-Menu

Provides for setting the cable loss and relative propagation velocity of the coax-
ial cable. Selected values may be changed using the Up/Down arrow key or key-
pad.

q

LOSS

— Opens the Cable Loss parameter for data entry. Enter the loss per

foot (or meter) for the type of transmission line being tested. Press

ENTER

when data entry is complete. (Range is 0.000 to 5.000 dB/m, 1.524 dB/ft)

q

PROP VEL

(relative propagation velocity) — Opens the Propagation Veloc-

ity parameter for data entry. Enter the propagation velocity for the type of
transmission line being tested. Press

ENTER

when data entry is complete.

(Range is 0.010 to 1.000)

q

CABLE

— Opens a common coaxial cable folder and custom coaxial cable

folder. Select either folder and use the Up/Down arrow key and

ENTER

to

make a selection. This feature provides a rapid means of setting both cable
loss and propagation velocity. (Refer to Appendix A for a listing of common
coaxial cables showing values for Relative Propagation Velocity and Nomi-
nal Attenuation in dB/m @ 1000 MHz, 2000 MHz and 2500 MHz.) The cus-
tom cable folder can consist of up to 24 user-defined cable parameters
downloaded via the Site Master Software Tools program.

q

WINDOW

— Opens a menu of FFT windowing types for the DTF calcula-

tion. Scroll the menu using the Up/Down arrow key and make a selection
with the

ENTER

key.

q

BACK

— Returns to the Distance Menu.

2 - 8

Chapter 2 Functions and Operations