Calibration Procedure - Single Port

Check that the “CAL OFF” message is displayed in the upper left corner of the display.
This indicates that the Site Master has not been calibrated. The following procedure details
how to perform the calibration.

Step 1. Select the appropriate frequency range, as described in the procedure on page

3-2.

Step 2. Press the

START CAL

key. The message

CONNECT OPEN TO RF Out

PORT will appear in the display.

Step 3. Connect the calibrated Open and press the

ENTER

key. The messages

Mea-

suring OPEN and then CONNECT SHORT TO RF Out PORT will appear.

Step 4. Remove the Open, connect the calibrated Short and press the

ENTER

key. The

messages

Measuring SHORT and then CONNECT TERMINATION TO RF

Out will appear.

Step 5. Verify that the calibration has been properly performed by checking that the

CAL ON message is now displayed in the upper left corner of the display.

3-3

Chapter 3 Getting Started

HOLD

RUN

START
CAL

AUTO
SCALE

SAVE
SETUP

RECALL
SETUP

LIMIT

MARKER

SAVE
DISPLAY

RECALL
DISPLAY

PRINT

MODE

FREQ/DIST

AMPLITUDE

SWEEP

SYS

ENTER

CLEAR

ESCAPE

ON

OFF

/

1

2

4

5

6

7

8

9

0

3

+

-

.

Site Master S332C

OPEN

LOAD

SHORT

CALIBRATION

RFOUT/REFLECTION

TEST PORT

TEST PORT CABLE (OPTIONAL)

Figure 3-2.

Calibration Setup

Advertising