Anritsu S251C User Manual

Page 54

Advertising
background image

Procedure - Return Loss Mode

Step 1. Press the

MODE

key.

Step 2. Select

FREQ-RETURN LOSS using the Up/Down arrow key and press

ENTER

.

Step 3. Set the start and stop frequencies, F1 and F2, as described on page 3-2.

Step 4. Connect the Test Port Extension cable to the RF Out port and calibrate the Site

Master as described on page 3-2.

Step 5. Save the calibration set up (page 3-5).

Step 6. Connect the Device Under Test to the Site Master phase stable Test Port Exten-

sion cable. A trace will be displayed on the screen as long as the Site Master is
in sweep mode.

Step 7. Press the

AMPLITUDE

key and set

TOP and BOTTOM values of the display.

In the example below, the TOP is set to 4, and the BOTTOM is set to 10.

Step 8. Press the

MARKER

key.

Step 9. Set

M1 to MARKER TO PEAK.

Step 10. Set

M2 to MARKER TO VALLEY.

Step 11. Calculate the insertion loss by averaging M1 (marker to peak) and M2 (marker

to valley) and dividing by two as follows:

Step 12.

Insertion Loss

M

M

=

+

1

2

2
2

=

Avg.

2

(where Avg. is one way loss)

Step 13. Press

SAVE DISPLAY

(page 3-5) name the trace, and press

ENTER

.

Figure 4-4 is an example of a typical insertion loss measurement in return loss mode.

Chapter 4 Measurements

4-7

Figure 4-4.

Typical Insertion Loss Measurement in Return Loss Mod