Anritsu S251C User Manual

Page 56

Advertising
background image

Figure 4-5 shows an example of a typical DTF return loss trace.

In the above example:

q

Marker M1 marks the first connector, the end of the Site Master phase stable Test
Port Extension cable. (M1 = 38.42 dB, 3.2 ft.)

q

Marker M2 marks the first jumper cable. (M2 = 32.04 dB, 13.1 ft.)

q

Marker M3 marks the end of the main feeder cable. (M3 = 31.70 dB, 149.0 ft.)

q

Marker M4 is the load at the end of the entire transmission line. (M4 = 43.10 dB,
157.0 ft.)

Procedure - DTF-SWR Mode

The following steps explain how to measure DTF in SWR mode.

Step 1. Press the

MODE

key.

Step 2. Select the

DTF-SWR using the Up/Down arrow key and press

ENTER

.

Step 3. Follow the same procedure as DTF-Return Loss Mode, above.

Chapter 4 Measurements

4-9

Figure 4-5.

Typical DTF Return Loss Trace