Typical Line Sweep Test Procedures

This section provides typical line sweep measurements used to analyze the performance of
a transmission feed line system.

System Return Loss Measurement

System return loss measurement verifies the performance of the transmission feed line sys-
tem with the antenna connected at the end of the transmission line. To measure the system
return loss:

Required Equipment

q

Site Master Model S251C

q

Precision Open/Short, Anritsu 22N50 or
Precision Open/Short/Load, Anritsu OSLN50LF

q

Precision Load, Anritsu SM/PL

q

Test Port Extension Cable, Anritsu 15NNF50-1.5C

q

Optional 510-90 Adapter, DC to 7.5 GHz, 50 ohm, 7/16 (F)-N (M)

Device Under Test

q

Transmission Feed Line with Antenna

Procedure

Step 1. Press the

MODE

key.

Step 2. Select

FREQ-RETURN LOSS using the Up/Down arrow key and press

ENTER

.

Step 3. Set the start and stop frequencies, F1 and F2, as described on page 3-2.

Step 4. Calibrate the Site Master as described on page 3-2.

Step 5. Connect the Device Under Test to the Site Master. A trace will be displayed on

the screen when the Site Master is in the sweep mode.

Step 6. Press

SAVE DISPLAY

(page 3-5) name the trace, and press

ENTER

.

NOTE: The antenna must be connected at the end of the transmission feed line
when conducting a System Return Loss measurement.

Chapter 4 Measurements

4-3