Device under test, Step 4. press the, Chapter 4 measurements – Anritsu S251C User Manual

Page 64: Mode, Enter, Freq/dist, Marker, Figure 4-13. tma-d measurement setup

Advertising
background image

Device Under Test

q

TMA-D

Procedure

Step 1. Connect the External Power Supply (40-115) to the Site Master.

Step 2. Turn on the Site Master and press the

MODE

key.

Step 3. Select

INSERTION GAIN (–30 dBm) using the Up/Down arrow key and press

ENTER

.

Step 4. Press the

FREQ/DIST

key to set the frequency limits. For a TMA-D amplifier,

set F1 to 1800 MHz and F2 to 1990 MHz.

Step 5. Perform a two-port calibration of the Site Master (page 3-4).

Step 6. Connect the RF Out port of the Site Master to the ANT connector of the TMA.

Step 7. Connect the RF In port of the Site Master to the RX connector of the TMA

Step 8. Press the

MARKER

key and set marker set M1 to 1834 MHz and marker M2 to

1871 MHz

Step 9. Press the

SYS

key, then

OPTIONS, then BIAS TEE to turn on the bias tee volt-

age. The gain is the level change before and after the bias tee voltage is turned
on.

4-17

Chapter 4 Measurements

TMA-D

RX TX ANT

HOLD

RUN

START
CAL

AUTO
SCALE

SAVE
SETUP

RECALL
SETUP

LIMIT

MARKER

SAVE
DISPLAY

RECALL
DISPLAY

PRINT

MODE

FREQ/DIST

AMPLITUDE

SWEEP

SYS

ENTER

CLEAR

ESCAPE

ON

OFF

/

1

2

4

5

6

7

8

9

0

3

+

-

.

Site Master S251C

RF IN TEST PORT

TEST PORT CABLE
(OPTIONAL)

TEST PORT CABLE

(OPTIONAL)

EXTERNAL POWER

SUPPLY (40-115)

RF OUT TEST PORT

(BIAS TEE)

Figure 4-13.

TMA-D Measurement Setup

Advertising