Chapter 3 _______________________________________________________________ Installation

VAISALA _______________________________________________________________________ 29

0505-274

Figure 11

Cleaning the Tightening Cone

HMP368 for Pressurized Pipelines
or Moisture in Liquids

Due to the sliding fit the HMP368 is easy to install into and remove
from a pressurized process. The probe is especially suitable for
measurements inside pipelines. See section

Installing the HMP368

Probe Through a Ball Valve Assembly on page 32

.

The following numbers refer to

Figure 11 on page 29

:

1

=

Fitting screw

2

=

Sealing washer

3

=

Tightening cone

4

=

Clean cotton stick

CAUTION

In pressurized processes it is essential to tighten the supporting nuts
and screws very carefully to prevent loosening of the probe by the
action of pressure.

NOTE

When HMP364 is installed in a process with a pressure differing from
normal atmospheric pressure, please enter the pressure value of the
process (in bar

a

) into the transmitter memory, see section

Setting the

Pressure for Calculations on page 48

.

CAUTION

In pressurized processes it is essential to tighten the supporting nuts
and screws very carefully to prevent loosening of the probe by the
action of pressure.