Calculating Correspondence of Current Values
and Output Quantities

When using HMT360 without a display, calibration and adjustment is
carried out by using a multimeter. Use the following equations to
calculate the current values corresponding to the reference output
quantities.

0605-015

Example:

Temperature scaling -40 ... +120 °C, reference 22.3 °C:

0605-017

WARNING

The power supply connectors (C1 and C2) MUST NOT be used in
hazardous areas. For calibration and adjustment in a hazardous area
use only an approved multimeter, which fulfills the safety factors
printed on the protective cover.

where

Q

ref

=

Reference value of the calibrated quantity

Q

min

=

Reference value of the calibrated quantity

Q

max

=

Value corresponding to 20 mA

I

4 mA

16 mA

Q

ref

Q

min

Q

max

Q

min

-----------------------------

+

=

I

4 mA

16 mA

22.3

° C 40° C

120

° C 40° C

--------------------------------------

+

10.230 mA

=

=