Adjustment commands, Ct temperature adjustment – Vaisala HMT360N User Manual
Page 57
Chapter 4 ________________________________________________________________ Operation
VAISALA _______________________________________________________________________ 55
Example:
>std -50 50<cr>
Td lo: -50.00 %RH
Td hi: 50.00 %RH
>
Adjustment Commands
Turn the internal DIP switch Calibration Enabled/Disabled to
position ON before the adjustment and return it to position OFF after
the adjustment.
CT Temperature Adjustment
Syntax: CT<cr>
The transmitter asks and measures temperature readings and calculates
the calibration coefficients.
Example:
>ct<cr>
T : 0.811 1. ref ? 0.5<cr>
Press any key when ready ...<cr>
T : 99.122 2. ref ? 99.5<cr>
OK
>
The OK indicates that the adjustment was succesful. When performing
one-point adjustment press only Enter for the second reference.