User's Guide ______________________________________________________________________

50 __________________________________________________________________ M210744EN-C

Lower Half of Display

The output quantities metric/non metric DIP switch can be used to
select between °C and °F.

It is possible to check the pressure setting of the transmitter by pressing
button Dn. To return to the temperature reading, press button Dn again.

Selecting Analog Outputs

You can select the output quantities for channels 1 and 2 by turning the
Calibration Enabled/Disabled DIP switch of the transmitter to
position Enabled (up) and the three output selection DIP switches to
position special (all up).

Press button C on the display cover: text "SCAL" appears on the
display. Then press button E until the following display appears:

0505-288

The characters on the second line of the display in this menu correspond
to the quantities according to the table in section

Selecting Output

Quantities on page 49

.

Select the quantity for Ch 1 with buttons Up and Dn and acknowledge
the selection with button E.

If the transmitter is equipped with two analog channels, select the
quantity for Ch 2 in the same way, for example:

A

water content

w

mg/m

3

lb/MMscf

Character

Quantity

Abbreviation

Metric Unit

Nonmetric Unit