Chapter 3 _______________________________________________________________ Installation

VAISALA _______________________________________________________________________ 31

0507-025

Figure 13

Sealing of Fitting Body into Process

Tightening the Clasp Nut

1.

Adjust the probe to a suitable depth according to the type of
installation.

2.

Tighten the clasp nut first manually.

3.

Mark the fitting screw and the clasp nut.

4.

Tighten the nut a further 50 ... 60° (ca. 1/6 turn) with a fork
spanner. If you have a suitable torque spanner, tighten the nut to 45
± 5 Nm (33 ± 4 ft-lbs).

CAUTION

Take care not to damage the probe body. A damaged body makes the
probe head less tight and may prevent it from going through the clasp
nut.