Serial communication settings, Setting the analog outputs, Asel selecting analog outputs – Vaisala HMT360N User Manual
Page 55: Table 6, Serial communications settings
Chapter 4 ________________________________________________________________ Operation
VAISALA _______________________________________________________________________ 53
Serial Communication Settings
To start issuing commands, make sure that HMT360 is connected to a
serial port of your computer and that the terminal session is open. Issue
commands by typing them on your computer according to the following
instructions. In these commands, <cr> stands for pressing Enter (on
your computer keyboard).
Setting the Analog Outputs
ASEL Selecting Analog Outputs
Syntax: ASEL [xxx yyy]<cr>
Table 6
Serial Communications Settings
Parameter
Value
bauds
2400
parity
none
data bits
8
stop bits
1
CAUTION
When using the serial interface, be sure that power supply and serial
interface are not connected to the same ground (use a floating power
supply or hand-held serial interface device).
where
xxx =
Quantity of Ch 1
yyy =
Quantity of Ch 2