4. selecting a weighing unit, 5. basic operation – A&D Ej Series Compact Balance EJ-2000 User Manual

Page 11

Advertising
background image

9

5-4. Selecting a weighing unit

Press the UNITS key to select a unit.

The following sections are a description of the
three common units: g (gram mode), pcs (counting
mode), and % (percent mode).






5-5. Basic operation

1. Select a weighing unit.

2. When the display does not show zero, press the

RE-ZERO key to set the display to zero.


3. When using a tare (container), place the

container on the weighing pan, and press the

RE-ZERO key to subtract the tare weight.

4. Place the object to be weighed on the pan or in

the container.
Wait for the STABLE indicator to be displayed
and read the value.


5. Remove the object from the pan.

‰

The RE-ZERO key will zero the balance if

the weight is within ±2% of the weighing
capacity around the power-on zero point.
The ZERO indicator turns on. When the
weight exceeds +2% of the weighing
capacity, it will be subtracted to zero as a
tare weight. In this case the ZERO and NET
marks turn on.

Precautions during operation

‰

Make sure that the STABLE indicator is on whenever reading or storing a

value.

‰

Do not press the keys with a sharp object such as a pencil.

‰

Do not apply a shock or a load to the pan that is beyond the weighing

capacity.

‰

Keep the balance free from foreign objects such as dust or liquid.

‰

Calibrate the balance periodically to keep weighing accuracy. (See “7.

CALIBRATION”.)


Each pressing switches the

units available in the order

described on the previous

page.

0.0

STABLE indicator

ZERO

Container

Weighing pan

0.0

NET

ZERO

150.3

NET

STABLE indicator

Advertising