9-6. EJ-13 Density Determination Kit

Using this option and calculation
program, the balance can determine
the density (specific gravity) of a
sample.

EJ-13 Installation

‰

Push the fixing frame out, and

remove the weighing pan.

‰

Attach the beaker stand to the balance and fit the

pan stand to the pan support of the balance.


























‰

Place a beaker filled with water on the beaker stand and place the sample weighing

pan on top of the pan stand.

Density (specific gravity) measurement

‰

The density of a liquid can be changed and there are two ways of setting. One is to

set the water temperature and the other is to set density value directly.

‰

The factory setting for density of a liquid is 25 °C as water temperature (the density

value ρ = 0.99704 (g/cm

3

) is used to calculate).

‰

The density (specific gravity) is calculated by the following formula.

S = X ρ

S: Density (specific gravity) of a sample

A: Weight in air

B: Weight in liquid

ρ: Density of liquid (water)

‰

The result is shown with two decimal places.

A

A-B

Weighing pan

Fixing frame

Sample weighing pan

Beaker

Pan stand

Upper pan to weigh in air

Lower pan to weigh in water

Pan support