Comparator, Poff 0 cp 0 cp 3, Func – A&D Ej Series Compact Balance EJ-2000 User Manual

Page 14: 1. setting example

Advertising
background image

12

6. COMPARATOR

The results of the comparison are indicated by HI, OK or LO on the display.
The comparison is as follows:
LO

< Lower limit value ≤ OK ≤ Upper limit value < HI

Operating conditions (see the function setting “Cp”):

‰

No comparison (comparator function disabled).

‰

Compares all data.

‰

Compares all stable data.

‰

Compares plus data except those near zero (plus data greater than +4d).

‰

Compares stable plus data except those near zero (stable plus data

greater than +4d).

‰

Compares all data except those near zero (all data greater than +4d or

less than -4d).

‰

Compares stable data except those near zero (stable data greater than

+4d or less than -4d).

d = the smallest display division. e.g.: 4d = four display divisions

The upper limit and lower limit numerical values are common to each of the weighing,
counting and percent mode. The example for EJ-120/200/300/410/610 is as follows.

Upper limit value “001010”: “10.10g” “1010pcs” “101.0%”
Lower limit value “000990”: “9.90g” “990pcs” “99.0%”


6-1. Setting example

This example will be “Compares plus data except
those near zero”.

Selecting a comparison mode

1. Press and hold the SAMPLE key to display

func .

(If the comparison mode is already set, press the

SAMPLE key to go to “Entering the upper and

lower limit values”.)


2. Press the PRINT key, then the balance displays

poff X .


3. Press the SAMPLE key several times to display

Cp X .


4. Press the RE-ZERO key several times to display

Cp 3 .


5. Press the PRINT key to store the settings.

Cp Hi appears after end .


poff 0

Cp 0

Cp 3

Press and hold

Press

several times

Press

several times

To store

func